ABS 1 Cu. Ft. Pharmacy Countertop Controlled Auto Defrost Freezer Freestanding
Our proprietary “Controlled Auto Defrost” technology provides the convenience of an auto-defrost design while still ensuring that vaccines remain at consistent and stable temperatures below 5°F (-15°C)
For more information on Vaccine Storage: Vaccine Storage
The acceptable temperature ranges for vaccine storage are between 36° Fahrenheit and 46° Fahrenheit (or between 2° Celsius and 8° Celsius) in the refrigerator and between -58° Fahrenheit and +5° Fahrenheit (or between -50° Celsius and -15° Celsius) in the freezer.
- Microprocessor Temperature Controller
- Digital Temperature Display With Sample Simulated Probe
- Audible And Visual High And Low Temperature Alarms
- Remote Alarm Contacts
- Operating Temperature Range: 5°F TO -13°F (-15°C TO -25°C)
- Defrost:
- Auto Defrost Cycle Every 6 Hours: (AT SET POINT -11°F /-24°C)
MAX AIR TEMP IN EMPTY FREEZER: 5°F (-15°C)
MAX AIR TEMP IN LOADED FREEZER: 3°F (-16°C)
MAX SAMPLE TEMP IN EMPTY FREEZER: 0°F (-18°C)